Unicel DXH Slidemaker Stainer
- Starting date:
- October 1, 2012
- Posting date:
- October 29, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15138
Recalled Products:
Unicel DXH Slidemaker Stainer
Reason
The DXH SMS systems generates an incorrect "invalid delimiters" message, when the host transmits the minimal header record. When this issue occurs, the DXH SMS does not store the test order in the worklist which results in the specimen being processed by a default order or skipped (when no default order is selected at the system manager).
Affected products
Unicel DXH Slidemaker Stainer
Lot or serial number
Software Version 1.5.1.0
Model or catalog number
775222
Companies
- Manufacturer
- Beckman Coulter Inc.