This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Unicel DXC 800 Pro and Unicel DXC 800I – Memory allocation error

Starting date:
April 9, 2012
Posting date:
May 14, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-26303

Recalled Products

A. Unicel DXC 800 Pro

B. Unicel DXC 800I

Reason

Memory allocation error (0x13890001): Accumulation of message manager tasks uses up available DXC Console Memory causing the Console to shut down.

Affected products

A. Unicel DXC 800 Pro

Lot or serial number
  • 2113
  • 2337
Model or catalog number

A11812

Companies
Manufacturer

Beckman Coulter, Inc.



B. Unicel DXC 800I

Lot or serial number

4143

Model or catalog number

A59102

Companies
Manufacturer

Beckman Coulter, Inc.