Health product recall

Ultraview DM3 Monitor

Last updated

Summary

Product
Ultraview DM3 Monitor
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Ultraview DM3 Monitor More than 10 numbers, contact manufacturer. 91330

Issue

The lithium-ion battery used in the DM3 is susceptible to failure and overheating when used beyond its recommended lifetime or otherwise damaged. This is due to a short in the battery with indeterminant cause, but likely attributed to use beyond service life, overcharging, or damage to the battery. Batteries known to be older than two years or batteries that are no longer holding a charge should be replaced. Damaged batteries must be replaced immediately.

Recall Start Date: September 27, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Zoe Medical, Incorporated

460 Boston Street, Topsfield, Massachusetts, United States, 01983

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76310