TUBERCULIN GLASS SYRINGES (2015-07-09)
- Starting date:
- July 9, 2015
- Posting date:
- August 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54740
Affected Products
A. Perfectum Tuberculin Glass Syringe
B. Micro-Mate Tuberculin Glass Syringe
Reason
Cadence Science, Inc. was notified by a customer that the graduation marks of the syringe are not centered with the numeric position marks on the syringe barrel. While the graduations marks are correctly spaced for volume, the numeric marks do not correctly align. This could cause the user to either under or over dose.
Affected products
A. Perfectum Tuberculin Glass Syringe
Lot or serial number
79713-04
W003542/1
W006324/3
Model or catalog number
5202
Companies
- Manufacturer
-
Cadence Science Inc
2080 PLAINFIELD PIKE
CRANSTON
02921
Rhode Island
UNITED STATES
B. Micro-Mate Tuberculin Glass Syringe
Lot or serial number
71880-01
80044-03
W002337/1
W002886/1
W006510/1
Model or catalog number
5208
5209
Companies
- Manufacturer
-
Cadence Science Inc
2080 PLAINFIELD PIKE
CRANSTON
02921
Rhode Island
UNITED STATES