TUBE EDTA PLH 13X75 3.0 PLBL LAV (2015-11-16)

Starting date:
November 16, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56202

Affected Products

TUBE EDTA PLH 13X75 3.0 PLBL LAV

Reason

The rubber stopper inside the Hemogard Cap has been dislodging after reinsertion of the stopper/cap subassembly onto the tube (referred to as "stopper creep" or "creep-out") or popping off of the tube which creates a potential for blood exposure and loss of sample.