This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TRX Vertical Positioner (2014-08-18)

Starting date:
August 18, 2014
Posting date:
October 2, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41649

Recalled Products

TRX Vertical Positioner 

Reason

Motion concept has identified through ongoing testing and as part of ongoing evaluation of the product return that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the vertical positioning system.

Affected products

TRX Vertical Positioner

Lot or serial number

30334

Model or catalog number

STRX-MPS-2

Companies
Manufacturer

Motion concepts IP,

84 Citation Drive, Units 1-7,

Concord,

L4K 3C1

Ontario

CANADA