Health product recall

Trutest Closed Suction System (2021-06-11)

Starting date:
June 11, 2021
Posting date:
June 17, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75827



Last updated:
2021-06-15

Affected Products

Trutest Closed Suction System

Reason

A swivel connector can become loose and may disconnect from ventilator circuit which would cause an interruption of ventilation to the patient. Disconnection is recognizable by the user as an alarm from the ventilator.

Affected products

Trutest Closed Suction System

Lot or serial number

KS2004020

Model or catalog number

T1972E114

Companies
Manufacturer

TRUDELL HEALTHCARE SOLUTIONS INC

758 BARANSWAY DRIVE

London

Ontario

CANADA