Trusight Cystic Fibrosis Library Prep Kit
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Trusight Cystic Fibrosis Library Prep Kit |
00816270020149 |
20036925 |
Issue
Annotation error in analysis software - the genomic coordinate reported for one variant in the sample did not match the cDNA name; the variant corresponding to the genomic coordinate is not cystic fibrosis-causing, but the cDNA name is potentially cystic fibrosis-causing (software issue/malfunction). Affected device details: catalog #20036925, affected software: local run manager (lrm) cf clinical seq 2.0 analysis module, software version:1.0.0.56, affected report file: trusightcfclinicalsequencingassay.txt.
Recall start date: May 25, 2022
Additional information
Details
Illumina, Inc.
5200 Illumina Way, San Diego, California, United States, 92122
Get notified