Trocar Catheter
- Starting date:
- October 26, 2016
- Posting date:
- November 25, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-61214
Reason
Atrium Trocar Catheters are packaged with a label containing an icon that shows more holes (eyelets) than the product actually contains. This graphical icon shows there are 6 side holes on the trocar catheter however, the correct number of eyelets on catheter is 2 side holes.
Affected products
Trocar Catheter
Lot or serial number
All lots
Model or catalog number
- 8408
- 8408
- 8410
- 8410
- 8412
- 8412
- 8416
- 8416
- 8420
- 8420
- 8424
- 8424
- 8428
- 8428
- 8432
- 8432