Health product recall

Triton 36 Ultrasonic Cleaner

Last updated

Summary

Product
Triton 36 Ultrasonic Cleaner
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Triton 36 Ultrasonic Cleaner

2307-XL36TR-055-1

6003000804

Issue

Our investigations have revealed a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the carbon fiber sensor over time. This degradation could potentially lead to water leakage into the machine.

Recall start date: April 1, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Ultra Clean Systems, Inc
281 Douglas Rd. E., Oldsmar, Florida, United States, 34677
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77476