Trima Accel Automated Blood Collection System, VE

Starting date:
May 1, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-63510

Reason

A Trima Accel device demonstrated a non-recoverable power failure. A subsequent investigation of the failure determined that a power filter cable, a component used in the internal electrical system, had a visible defect which caused the unit to lose power. Further evaluation determined that this defect may be present in other electrical cables which could cause a similar non-recoverable power failure.

Affected products

Trima Accel Automated Blood Collection System, VE

Lot or serial number

1T07708-1T08066

Model or catalog number

81000

Companies

Manufacturer

Terumo BCT Inc.

10811 W. Collins Avenue

Lakewood

Colorado

UNITED STATES