This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Trima Accel Automated Blood Collection System, VE

Starting date:
May 1, 2017
Posting date:
June 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-63510

Reason

A Trima Accel device demonstrated a non-recoverable power failure. A subsequent investigation of the failure determined that a power filter cable, a component used in the internal electrical system, had a visible defect which caused the unit to lose power. Further evaluation determined that this defect may be present in other electrical cables which could cause a similar non-recoverable power failure.

Affected products

Trima Accel Automated Blood Collection System, VE

Lot or serial number

1T07708-1T08066

Model or catalog number

81000

Companies

Manufacturer
Terumo BCT Inc.
10811 W. Collins Avenue
Lakewood
Colorado
UNITED STATES