This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Trilogy 200 Ventilators

Starting date:
May 15, 2012
Posting date:
June 11, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-26327

Recalled Products

  1. Trilogy 200 Ventilators

Reason

Units shipped contain a potentially defective component which could affect the power supply of Trilogy Ventilator. Should this problem occur there could be a loss of alarm function in conjunction with loss or reduction in therapy.

Affected products

A. Trilogy 200 Ventilators 

Lot or serial number
  • Made in March 2012
Model or catalog number
  • CA1032800
Companies
Manufacturer

Respironics Inc.