This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Trigen Titanium Nail System - Intertan Short Nails (2014-04-07)

Starting date:
April 7, 2014
Posting date:
June 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39803

Recalled products

Trigen Titanium Nail System - Intertan Short Nails

Reason

Smith & Nephew has initiated a product recall of several batches of the Trigen Intertan nail due to a manufacturing error, which resulted in the set screw being set too deep in the affected nails. During implantation, when the proximal femur is being reamed to accept the lag and compression screw, the reamer may impinge the set screw, causing potential breakage of the reamer.

Affected products

Trigen Titanium Nail System - Intertan Short Nails

Lot or serial number
  • 13LT36306
  • 13LT36308
  • 13LT36313
  • 13LT36398
  • 13LT37010
  • 13LT37012
Model or catalog number
  • 71675201
  • 71675202
  • 71675204
  • 71675205
  • 71675211
Companies
Manufacturer

Smith and Nephew, Inc.

1450 Brooks Road

Memphis

38116

Tennessee

UNITED STATES