This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TRIATHLON MODULAR HANDLE (2016-06-29)

Starting date:
June 29, 2016
Posting date:
July 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59410

Affected products

A. TRIATHLON MODULAR HANDLE

Reason

Analysis of two instruments from finished goods indicated that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved by the supplier, potentially resulting in disassociation from the instrument.

Affected products

A. TRIATHLON MODULAR HANDLE

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6541-4-808

Companies
Manufacturer

Howmedica Osteonics Corporation

325 Corporate Drive

Mahwah

New Jersey

UNITED STATES