Triathlon Distal Capture Assembly (2015-02-12)
- Starting date:
- February 12, 2015
- Posting date:
- March 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52599
Recalled Products
A) Triathlon Distal Capture Assembly
Reason
A material analysis determined that the three (3) returned PER (product experience report) devices showed no evidence of a line to line or interference fit between the cross pin and action trigger subcomponents as required by the design.
Affected products
A) Triathlon Distal Capture Assembly
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
6541-1-723
Companies
- Manufacturer
-
Stryker Orthopeadics
325 Corporate Drive
Mahwah
New Jersey
UNITED STATES