Transport Strut Hoffmann Lrf
Brand(s)
Last updated
Summary
Product
Transport Strut Hoffmann Lrf
Issue
Medical devices - Performance
What to do
Contact manufacturer if you require additional information
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Transport Strut Hoffmann Lrf | H30794 H32791 | 49330400 |
Issue
Stryker has identified a nonconformance in specific lots of Hoffmann Lrf bone transport struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked between 0- 50mm of the threaded rod.
Recall start date: February 6, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Stryker Gmbh
Bohnackerweg 1, Selzach, Solothurn, Switzerland, 2545
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72685
Get notified
Receive notifications for new and updated recalls and alerts by category.