Health product recall

Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long)

Last updated

Summary

Product
Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products Lot or serial number Model or catalog number
Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long) Lot H23J23039 and higher R5C4484

Issue

Baxter Healthcare Corporation is issuing a safety alert for the Minicap Extended Life Pd Transfer Sets listed due to an increase in complaints related to separation of the connector or spike from the main body of the transfer set. These separations may occur while connecting or disconnecting from the patient line, however as the tubing remains connected to the connector or spike during a separation, the sterile fluid path is maintained and there is no risk of microbial contamination that will result from the separation itself. 

Recall Start Date: October 21, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Baxter Healthcare Sa

Thurgauerstrasse 130, Glattpark (Opfikon), Switzerland, 8152

Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type II
Identification number
RA-76426

Get notified

Receive notifications for new and updated recalls and alerts by category.