Health product recall

Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long)

Last updated

Summary

Product
Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalog number

Transfer Set (Minicap Extended Life Pd Transfer Set With Twist Clamp – Extra Long)

Lot H20B11116 and higher

R5C4484

Issue

Baxter Healthcare Corporation (Baxter) is issuing an urgent medical device correction for the Minicap Extended Life Pd Transfer Sets listed, which are manufactured with peroxide-cured silicone tubing that transfers solution to and from the patient.

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to nondioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices. The source of the NDL PCBAs and/or NDL PCBs in those recalls was due to the manufacturing process of the silicone tubing.

Recall Start Date: October 21, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies

Baxter Healthcare Sa

Thurgauerstrasse 130, Glattpark (Opfikon), Switzerland, 8152

Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type II
Identification number
RA-76427

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe