Trackmaster Treadmills
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Trackmaster Treadmills |
Contact manufacturer. |
TMX428 220 |
Issue
This involves the motor/drive pcb combination of the GE Healthcare T2100-ST series and the Trackmaster series treadmills. The motor/drive pcb configuration has inherent noise between the encoder and the drive pcb which may cause the drive pcb to misread the orientation of the motor when commanded to start movement. This belt may accelerate forwards or backwards without notice.
Recall Start Date: May 24, 2023
Additional information
Details
Full Vision Inc
3017 Full Vision Dr., Newton, Kansas, United States, 67114
Get notified