Health product recall

Tornier Perform® Humeral Fracture Locking Screwdriver

Brand(s)
Last updated

Summary

Product
Tornier Perform® Humeral Fracture Locking Screwdriver
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number/

Tornier Perform® Humeral Fracture Locking Screwdriver

22A925

IFX814

Issue

Stryker has identified a nonconformance in one specific lot of Tornier Perform Humeral Fracture Locking Screwdriver. Specifically, the screwdriver is not able to retain screws as expected. The screwdriver's main function (tightening locking screws) during surgery is still able to function, only the ability to retain the screw on the driver is impacted.

Recall start date: May 13, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

 

Tornier Inc.

10801 Nesbitt Avenue South, Bloomington, Minnesota, United States, 55437

 

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75555

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe