Toothed Rack Retractor for Matrix and Removal Tool for T-PAL (2013-10-25)

Starting date:
October 25, 2013
Posting date:
December 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36977

Recalled products

  1. Toothed Rack Retractor, for Matrix
  2. Removal Tool for T-PAL

Reason

Synthes was notified by its supplier that within one material batch used for production of the affected parts one or several bars were mixed in with an incorrect material. Specified material was 420A and material of the mixed in bar(s) was X20CRMNNI 12 8 6 (which is a non-magnetic austenitic steel grade used for electric generators). The wrong material results in lower than specified hardness and corrosion resistance.

Affected products

A. Toothed Rack Retractor, for Matrix

 

Lot or serial number
  • T987270
Model or catalog number
  • 03.632.087
Companies
Manufacturer

Synthes (Canada) Ltd.

2566 Meadowpine Blvd.

Mississauga

L5N 6P9

Ontario

CANADA



B. Removal Tool for T-PAL

Lot or serial number
  • T987623
Model or catalog number
  • 03.812.005
Companies
Manufacturer

Synthes (Canada) Ltd.

2566 Meadowpine Blvd.

Mississauga

L5N 6P9

Ontario

CANADA