Health product recall

Titan Surgical Table (2014-10-17)

Starting date:
October 17, 2014
Posting date:
November 7, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42033

Recalled Products

Titan Surgical Table

Reason

Manufacturer has become aware of unintended movement occurring during surgical cases due to the use of damaged titan remotes. There is a possibility that damage can occur to a remote control, which breaks the button, and can cause permanent or sporadic electrical conduction and trigger unintended movement. This is a rare event and can be avoided by proper handling and maintenance of the remote control. The product is safe as long as it is not damaged.

Affected products

Titan Surgical Table

Lot or serial number

All lots.

Model or catalog number

1228073

Companies
Manufacturer
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY