Titan Surgical Table (2014-10-17)
- Starting date:
- October 17, 2014
- Posting date:
- November 7, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-42033
Recalled Products
Titan Surgical Table
Reason
Manufacturer has become aware of unintended movement occurring during surgical cases due to the use of damaged titan remotes. There is a possibility that damage can occur to a remote control, which breaks the button, and can cause permanent or sporadic electrical conduction and trigger unintended movement. This is a rare event and can be avoided by proper handling and maintenance of the remote control. The product is safe as long as it is not damaged.
Affected products
Titan Surgical Table
Lot or serial number
All lots.
Model or catalog number
1228073
Companies
- Manufacturer
-
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY