Health product recall

Titan Blis-Sonic SW Scaler

Last updated

Summary

Product
Titan Blis-Sonic SW Scaler
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Titan Blis-Sonic SW Scaler 4874201 261669 264563

Issue

A voluntary recall was initiated for a certain lot of Titan Scaler Tip-Perio (part number# 261669) specifically packaged with Blis-Sonic Scaler (part number# 264563) lot# 4874201 distributed between March 06, 2024 through May 10, 2024. This recall is initiated because it was discovered during production testing that the tip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder.

Recall Start Date: June 4, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies

Dentalez, Inc., Stardental Division

1816 Colonial Village Lane, Lancaster, Pennsylvania, United States, 17601

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75825

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe