This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Tina-quant Cystatin C (2015-05-14)

Starting date:
May 14, 2015
Posting date:
June 15, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53791

Recalled Products

  1. Tina-quant Cystatin C
  2. Tina-quant Cystatin C (CYSC)
  3. CYSC (Tina-quant Cystatin C)

Reason

The specification for "high dose hook effect" for the CYSC2 on all Roche systems has been revised and will be reduced from 20 mg/l to 12 mg/l.

Affected products

A. Tina-quant Cystatin C

Lot or serial number

All lots.

Model or catalog number
  • 6600174190
Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. Tina-quant Cystatin C (CYSC)

Lot or serial number

All lots.

Model or catalog number
  • 6600239190
Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY



C. CYSC (Tina-quant Cystatin C)

Lot or serial number

All lots.

Model or catalog number
  • 6600263190
Companies
Manufacturer

Roche Diagnostics Gmbh

Sandhoferstrasse 116

Mannheim

68305

GERMANY