Thoracic Grasper (2015-05-18)
- Starting date:
- May 18, 2015
- Posting date:
- October 14, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55330
Affected Products
Thoracic Grasper
Reason
Coating of 5mm Thoracic Grasper main tube may experience damage. Worst case effect of this damage is the main tube coating separating from the instrument and depositing fragments of material into the patient.
Affected products
Thoracic Grasper
Lot or serial number
- A10120120
- A10120120
- S10110617
- S10110617
- S10120515
- S10120515
- S10130408
- S10130408
- S10130613
- S10130613
- S10131004
- S10131004
- S10131022
- S10131022
Model or catalog number
- 420343
- 420343
Companies
- Manufacturer
-
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
California
UNITED STATES