ThermoScientific Oxoid Columbia Blood Agar Base
- Starting date:
- August 11, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-60364
Affected Products
Reason
A technical investigation has identified that certain lots of Oxoid Columbia Blood Agar Base, CM0331, may not provide acceptable microbiological performance when used in selective formulations targeting staphylococcus and streptococcus species. However, these formulations continue to perform correctly with the quality control (QC) strains detailed in the QC specs. Performance with some clinical strains may be atypical.
Affected products
ThermoScientific Oxoid Columbia Blood Agar Base
Lot or serial number
1841146
1857212
Model or catalog number
CM0331
Companies
- Manufacturer
-
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM