Thermo Scientific Oxoid Mueller Hinton Broth (2018-03-07)
- Starting date:
- March 7, 2018
- Posting date:
- April 6, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66390
Affected products
Thermo Scientific Oxoid Mueller Hinton Broth
Reason
An internal technical investigation by Oxoid Limited, part of Thermo Fisher Scientific, has confirmed that Thermoscientific Oxoid Mueller Hinton Broth, CM0405B lot 1765821 has incorrect magnesium cation information detailed on the product label.
Current values: 2.293mg mg++ per litre
Correct values: 4.883mg mg++ per litre
Affected products
Thermo Scientific Oxoid Mueller Hinton Broth
Lot or serial number
1765821
Model or catalog number
CM0405B