TFL Fiber Cleaver and Stripper
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
TFL Fiber Cleaver |
All lots. |
TFL-AFC |
TFL Fiber Stripper |
All lots. |
TFL-AFS550 |
Issue
Olympus is taking this medical device recall action after a review of reprocessing instructions for the Fiber Strippers and Cleavers. Both the strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
Recall Start Date: December 11, 2023
Additional information
Details
Gyrus Acmi, Inc
9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445