Tensive Conductive Adhesive Gel
Brand(s)
Last updated
Summary
Product
Tensive Conductive Adhesive Gel
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Tensive Conductive Adhesive Gel |
A0623009 |
22-60 |
Issue
Parker has received five complaints related to the viscosity of tensive conductive adhesive gel. Customers reported the gel is too thick, darker than usual, and in one case patient irritation.
Recall start date: January 20, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Parker Laboratories Inc.
286 Eldridge Road, Fairfield, New Jersey, United States, 07004
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75047
Get notified