This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Tempspan CMT (N69CA)

Starting date:
August 22, 2012
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-15557

Recalled Products

A. Tempspan CMT

Reason

The direction for use specify a 60-second working time which is not indicative of the actual performance of the product. Data shows that the product has a minimum working time of 45 seconds.

Affected products

A. Tempspan CMT

Lot or serial number

All Lots

Model or catalog number

N69CA

Companies
Manufacturer
Kerr Corporation