This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Temporary Myocardial Bipolar Heartwire

Starting date:
July 25, 2012
Posting date:
November 26, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15888

Recalled Products

A. Temporary Myocardial Bipolar Heartwire

Reason

Spacing between zig-zag wire and insulation is too long (10mm instead of 2mm). The defect affects the proper placement of zig-zag into the heart wall and subsequent electrical contact might be compromised.

Affected products

A. Temporary Myocardial Bipolar Heartwire

Lot or serial number

C1-07056

Model or catalog number

TME 64-S

Companies
Manufacturer

Oscor Inc.