This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TELESENTRY BATTERY (2016-03-08)

Starting date:
March 8, 2016
Posting date:
April 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57752

Affected Products

TELESENTRY BATTERY

Reason

Aging batteries may become hot and cause burns to user due to a potential short (thermal runaway).  The batteries should be replaced after 2 years or 300 charge cycles per the IFU however no manufacturing date is printed on the battery so it is difficult for the user to gauge correctly.

Affected products

TELESENTRY BATTERY

Lot or serial number

All

Model or catalog number

101096

Companies
Manufacturer
Scottcare Corporation
4791 W. 150TH STREET
CLEVELAND
44135
Ohio
UNITED STATES