TDHistoCyto Software
Brand(s)
Last updated
Summary
Product
TDHistoCyto Software
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
TDHistoCyto Software | Version 13.61 and later | TD-TDH-01 |
Issue
A software bug was reported by a user site. The end user noticed a discrepancy between the values of a number of additional data displayed in MS Word format compared with what is printed, including what is displayed in the PDF report after validation. When the problem occurs, the PDF report that can be consulted by prescribers does not include the latest updated values of additional data.
Recall start date: October 15, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Technidata S.A.S.
387, Avenue Jean Kuntzmann, Montbonnot Saint Martin, France, 38330
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76371
Get notified