Health product recall

TDHistoCyto Software

Last updated

Summary

Product
TDHistoCyto Software
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

TDHistoCyto Software

Software version V13.21.B

TD-TDH-01

Issue

Inappropriate behaviour was noted on a TDHC user site, specifically with regard to send/receipt by an external host system. In some rare cases, analysis reports were included in the hospital information system (HIS) with results from another patient. This recall applies only to systems with an INST0031 connection.

Recall Start Date: August 28, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Technidata S.A.S.

387, Avenue Jean Kuntzmann, Montbonnot Saint Martin, France, 38330

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76595

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe