Health product recall

TCM5 monitor

Last updated

Summary

Product
TCM5 monitor
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Tcm5

software versions 1.5.0
software versions 1.5.0.1

393-500

Issue

Radiometer has become aware of an issue with the TCM5 monitor.
The issue is that the TCM5 application may freeze and not update the display with new values or send new values to a central monitoring system. This means that the last measured values remain displayed unchanged on the screen of the TCM5 screen and in a central monitoring system and may not reflect patient status.
For this reason, Radiometer requests that you stop using the affected product (except in sleep studies*), with immediate effect.
*All TCM5 monitors with software versions 1.5.0 or 1.5.0.1 are affected. However, it has been assessed that if used in sleep studies, non-critical patients will not experience harm.

Recall start date: October 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Radiometer Medical Aps
Akandevej 21, Bronshoj, Denmark, 2700
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76454

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe