This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TCM COMBI M MODULES (2015-11-16)

Starting date:
November 16, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56288

Affected Products

TCM COMBI M MODULES (2015-11-16)

Reason

After running without any problems and with the volume control set to 1 the analyzer gave an alarm which was significantly louder than the specified highest possible level.

Affected products

TCM COMBI M MODULES (2015-11-16)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

903-111

Companies
Manufacturer

Radiometer Medical Aps

Akandevej 21

Bronshoj

2700

DENMARK