This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
TARGET 360 NANO DETACHABLE COIL (2016-06-08)
- Starting date:
- June 8, 2016
- Posting date:
- June 24, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, Hospitals, General Public
- Identification number:
- RA-59030
Affected products
- TARGET 360 NANO DETACHABLE COIL
- TARGET 360 NANO DETACHABLE COIL
Reason
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Affected products
A. TARGET 360 NANO DETACHABLE COIL
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- M0035421020
- M0035421030
- M0035421520
- M0035421530
- M0035421540
- M0035442030
- M0035442040
- M0035442540
- M0035443040
- M0035443060
- M0035443560
Companies
- Manufacturer
-
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES
B. TARGET 360 NANO DETACHABLE COIL
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- M0035431010
- M0035431020
- M0035431030
- M0035431510
- M0035431520
- M0035431530
- M0035431540
- M0035452030
- M0035452040
- M0035452060
- M0035452530
- M0035452540
- M0035452560
- M0035453060
Companies
- Manufacturer
-
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES