This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TARGET 360 NANO DETACHABLE COIL (2016-06-08)

Starting date:
June 8, 2016
Posting date:
June 24, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, Hospitals, General Public
Identification number:
RA-59030

Affected products

  1. TARGET 360 NANO DETACHABLE COIL
  2. TARGET 360 NANO DETACHABLE COIL

Reason

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Affected products

A. TARGET 360 NANO DETACHABLE COIL

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • M0035421020
  • M0035421030
  • M0035421520
  • M0035421530
  • M0035421540
  • M0035442030
  • M0035442040
  • M0035442540
  • M0035443040
  • M0035443060
  • M0035443560
Companies
Manufacturer
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES

B. TARGET 360 NANO DETACHABLE COIL

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • M0035431010
  • M0035431020
  • M0035431030
  • M0035431510
  • M0035431520
  • M0035431530
  • M0035431540
  • M0035452030
  • M0035452040
  • M0035452060
  • M0035452530
  • M0035452540
  • M0035452560
  • M0035453060
Companies
Manufacturer
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES