This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND (2015-06-18)
- Starting date:
- June 18, 2015
- Posting date:
- July 17, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54220
Affected Products
A. TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND
Reason
The above mentioned product lot contains a weak Anti-Wr(a). In cells carrying a Wr(a) Antigen, the product may give a weak taise-positive reaction.
Affected products
A. TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND
Lot or serial number
8416100-01
Model or catalog number
806 530 100
Companies
- Manufacturer
-
BioRad Medical Diagnostics GMBH
INDUSTRIESTR.1
DREIEICH
63303
GERMANY