This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND (2015-06-18)

Starting date:
June 18, 2015
Posting date:
July 17, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54220

Affected Products

A. TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND

Reason

The above mentioned product lot contains a weak Anti-Wr(a). In cells carrying a Wr(a) Antigen, the product may give a weak taise-positive reaction.

Affected products

A. TANGO OPTIMO AUTOMATED BLOOD BANK ANALYZER SYS- SOLIDSCREEN II ANTI-D BLEND

Lot or serial number

8416100-01

Model or catalog number

806 530 100

Companies
Manufacturer

BioRad Medical Diagnostics GMBH

INDUSTRIESTR.1

DREIEICH

63303

GERMANY