This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity (2016-07-18)

Starting date:
August 18, 2016
Posting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59966

Affected Products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity

Reason

A microplate was inadvertantly placed in the loading station while in function, displacing the previous plate. A erroneous test result was attributed to the placement of the wrong microplate.

Affected products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity

Lot or serial number

All lots.

Model or catalog number

850000010

Companies
Manufacturer

Bio-Rad Medical Diagnostics GMBH

Industriestr.1

Dreieich