Health product recall

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity (2016-07-18)

Starting date:
August 18, 2016
Posting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59966

Affected Products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity

Reason

A microplate was inadvertantly placed in the loading station while in function, displacing the previous plate. A erroneous test result was attributed to the placement of the wrong microplate.

Affected products

TANGO AUTOMATED BLOOD BANK ANALYZER SYSTEM - Tango Infinity

Lot or serial number

All lots.

Model or catalog number

850000010

Companies
Manufacturer
Bio-Rad Medical Diagnostics GMBH
Industriestr.1
Dreieich