Health product recall

TactiSys Quartz Equipment (2019-04-24)

Starting date:
April 24, 2019
Posting date:
May 10, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69884


Last updated:

Affected Products

TactiSys Quartz Equipment

Reason

Device operating on software v1.7.0 may display intermittent contact force data during procedures. The device log data fills the allocated disk space, which prevents the storage on new log data. In such instances, the contact force information is displayed in purple indicating that the data is not current and may not be accurate. This data will remain purple until communication is re-established or the disk has room to store log files.

Affected products

TactiSys Quartz Equipment

Lot or serial number

Not applicable.

Model or catalog number

PN-004 400

Companies
Manufacturer

St. Jude Medical, Inc

One St. Jude Medical Drive

Saint Paul

55117

Minnesota

UNITED STATES