Health product recall

T2100 Treadmill

Last updated

Summary

Product
T2100 Treadmill
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
T2100 Treadmill GEDC-6676 GEDC-6823 GEDC-6622 GEDC-6611 GEDC-6610 GEDC-6751 2097357-002 2097357-001

Issue

Full vision is conducting a corrective action related to the magnetic safety tether of the GE Healthcare T2100-ST Treadmill. When the magnetic safety tether is pulled, the treadmill may fail to stop resulting in a potential injury.

Recall start date: November 14, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies

Ge Medical Systems Information Technologies Inc.

9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72211

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe