T2100 Treadmill (2015-02-12)
- Starting date:
- February 12, 2015
- Posting date:
- March 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52581
Recalled products
- T2100 Treadmill
Reason
Manufacturer internal quality inspection has found that the power cord connecting directly to the T2100 treadmill may not have been assembled according to specifications. If the power cord was improperly assembled and a separate secondary electrical fault condition exists (e.g., a frayed extension or power cord touching the treadmill chassis), this could possibly result in an electrical shock to the patient or operator. There have been no reported incidences of the treadmill power cord leading to an electrical shock of a patient or operator at this time.
Affected products
A. T2100 Treadmill
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- T2100
Companies
- Manufacturer
-
GE Medical Systems Information Technologies
Inc.
8200 West Tower Avenue
Wisconsin
UNITED STATES