This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

T2100 Treadmill (2015-02-12)

Starting date:
February 12, 2015
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52581

Recalled products

  1. T2100 Treadmill

Reason

Manufacturer internal quality inspection has found that the power cord connecting directly to the T2100 treadmill may not have been assembled according to specifications. If the power cord was improperly assembled and a separate secondary electrical fault condition exists (e.g., a frayed extension or power cord touching the treadmill chassis), this could possibly result in an electrical shock to the patient or operator. There have been no reported incidences of the treadmill power cord leading to an electrical shock of a patient or operator at this time.

Affected products

A. T2100 Treadmill

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • T2100
Companies
Manufacturer

GE Medical Systems Information Technologies

Inc.

8200 West Tower Avenue

Wisconsin

UNITED STATES