This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

T1 Transport Monitor/Module Battery Pack

Starting date:
May 28, 2017
Posting date:
June 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63620

Reason

MindRay has received two (2) reports that the battery pack for the TL transport monitor /module overheated forming a small, melted cavity in the outer housing. In both reports, the battery emitted smoke and was hot to the touch. In one (1) report a small flame was noted on the battery surface.

Affected products

T1 Transport Monitor/Module Battery Pack

Lot or serial number

  • DS 10 X XX XXXXX
  • DS 17 E XX XXXXX
  • DS 17 F XX XXXXX
  • DS 17 G 04 XXXXX
  • DS 17 G 08 XXXXX

Model or catalog number

Not Applicable

Companies

Manufacturer

Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Mindray Building, Suite 3B, Keji 12th Road South, Hi-Tech Industrial Park

Nanshan, Guangdong

518057

CHINA