Sysmex DI-60™ Main Unit
Brand(s)
Last updated
Summary
Product
Sysmex DI-60™ Main Unit
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Sysmex DI-60™ Main Unit |
All lots. |
CC286297 |
Issue
When analyzing a slide with unreadable barcode in a DI-60, there is a risk of incorrect barcode reading and misattribution of results. This risk is higher with certain readers and barcode placements. Interim measures are advised as per notice, and firmware upgrades will be provided to affected units.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Sysmex Corporation
4-4-4- Takatsukadai, Kobe, Japan, 651-2271
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78384
Get notified
Receive emails about new and updated recall and safety alerts.