Health product recall

Syngo.plaza (2020-08-14)



Starting date:
August 14, 2020
Posting date:
September 18, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73919


Last updated:

Affected Products

Syngo.plaza

Reason

Siemens is releasing a software update VB30C_HF03 that provides feature improvements and resolution of software errors including the ones listed below:

  • The syngo.plaza viewer could crash when studies with doubled private tags were received;
  • Sporadic error message that a study could not be opened in the syngo.plaza viewer;
  • Certain images from specific modalities could lead to a crash of the viewer;
  • Under certain circumstances, when the current study had "smart read data" only, prior data was shown instead of the current data;
  • Under special circumstances, images of different patients were shown under one patient tab.

Affected products

Syngo.plaza

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 11064436
  • 11249920
  • 11293724
Companies
Manufacturer

Siemens Healthcare GMBH

Henkestr. 127

Erlangen

91052

GERMANY