Syngo.Plaza (2020-02-12)
- Starting date:
- February 12, 2020
- Posting date:
- March 6, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-72493
Last updated: 2020-03-10
Affected Products
Syngo.Plaza
Reason
Siemens is releasing a software update VA30C_HF02 that resolves an error of a missing clause in SQL statement that causes unintended deletion/de-linking of Presentation State (PR) series from the Long Term Storage (LTS) archive database. The issue may occur in scenario's where there are no other image modality series in LTS under the study except the one currently being deleted from patient browser.
Affected products
Syngo.Plaza
Lot or serial number
100628
Model or catalog number
11064436
Companies
- Manufacturer
-
Siemens Healthcare GMBH
Henkestr. 127
Erlangen
91052
GERMANY