SYNGO.PLAZA (2018-11-11)
- Starting date:
- November 11, 2018
- Posting date:
- November 30, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68510
Affected products
SYNGO.PLAZA
Reason
Siemens Healthcare is notifying customers of syngo.plaza (version vb30a_hf01 and previous versions) that there is potential for the image orientation labels to be displayed incorrectly. The issue happens once images have been loaded and the user performs rotate and flip operations together and saved changes. The error scenario does not happen when only one of either rotate or flip is performed.
Affected products
SYNGO.PLAZA
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
11064436
11249920
11293724