SYNGO.PLAZA
Brand(s)
Summary
Contact the manufacturer if you require additional information
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
SYNGO.PLAZA |
More than 10 numbers, contact manufacturer. |
11293724 |
Issue
MRI images with certain values in the Dicom tags rescale slope (0028,1053) and rescale intercept (0028, 1052) can show wrong statistical values 'min/max' (maximum and minimum grayscale value in the ROI) and 'Mean/STD D' (mean value and standard deviation of the grayscale values) in the ROI, which can lead to an inconsistent, conflicting holistic view on the patient Recall start date: 2021-09-30 |
Additional information
Details
SIEMENS HEALTHCARE LIMITED
1577 NORTH SERVICE ROAD EAST, OAKVILLE
ON, CANADA
L6H 0H6