This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SYNGO X WORKPLACE WORKSTATION AND SOFTWARE (2015-11-17)

Starting date:
November 17, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56190

Affected Products

  1. SYNGO X WORKPLACE-WORKSTATION AND COMPUTER HARDWARE
  2. SYNGO X WORKPLACE-SOFTWARE

Reason

During regular testing at Siemens it was found that incorrect measurements in multi-planar reformattings (MPRS) of a volume were acquired with a CT scanner with a titled gantry.  In this situation, the "4d" tab card will incorrectly display length measurements that are too small, if the measurement has been performed at the upper right MPR in a 2x2 layout.

Affected products

A. SYNGO X WORKPLACE-WORKSTATION AND COMPUTER HARDWARE

Lot or serial number

12537

Model or catalog number

10281299

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



B. SYNGO X WORKPLACE-SOFTWARE

Lot or serial number

20060

20061

Model or catalog number

10502522

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY