SYNGO X WORKPLACE WORKSTATION AND SOFTWARE (2015-11-17)
- Starting date:
- November 17, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56190
Affected Products
- SYNGO X WORKPLACE-WORKSTATION AND COMPUTER HARDWARE
- SYNGO X WORKPLACE-SOFTWARE
Reason
During regular testing at Siemens it was found that incorrect measurements in multi-planar reformattings (MPRS) of a volume were acquired with a CT scanner with a titled gantry. In this situation, the "4d" tab card will incorrectly display length measurements that are too small, if the measurement has been performed at the upper right MPR in a 2x2 layout.
Affected products
A. SYNGO X WORKPLACE-WORKSTATION AND COMPUTER HARDWARE
Lot or serial number
12537
Model or catalog number
10281299
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. SYNGO X WORKPLACE-SOFTWARE
Lot or serial number
20060
20061
Model or catalog number
10502522
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY