Syngo Imaging - System Luk (2014-04-17)
- Starting date:
- April 17, 2014
- Posting date:
- May 5, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-39321
Recalled Products
Syngo Imaging - System Luk
Reason
Siemens has been informed from one customer that the cleanup procedure which is enabled with archving method "explicit archiving" and intended to clean up the working directory, may by mistake delete data from the archive itself in case the following two conditions are met:
- The configured path for the working directory is directly linked to the archive location
- The long term storage device does not have an appropraite deletion protection policy enabled
Affected products
Syngo Imaging - System Luk
Lot or serial number
1200
1262
1457
Model or catalog number
10 014 063
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY